US508MT Digital Flow Meter

US208MTUser Manual-V20240127The flow rate displayed is in litres per minute. The total flow displayed is in Litres. (By pressing SHIFT to shift the display)How to setProvide5-24V DCpower to light on,

  • Model: US508MT

US508MT User Manual-V20240127

 

The flow rate displayed is in litres per minute. The total flow displayed is in Litres. (By pressing SHIFT to shift the display)

How to set

Provide 5-12V DC power to light on.

Open the back cover of the device

Press SET for 5 seconds to enter the menu or parameters

Then press SET to select the menu

Press + or - to increase or decrease the number, hold + or – to change the numbers quickly.

Parameters 

1.       1.00500-K value, this setting value is 100 times the real K value

For example: the formula of the flow sensor is F=8Q, then real K=8, setting K is 800;

Range: 1-65535

Default value:500

2.       2.00500 – upper limit alarm value

=500L/Min means the alarm will go off when the flow rate is greater than the alarm value

Range: 1-65535

Default value:500

3.       3.001.00 – lower limit alarm value

=1.00 means the alarm will go off when the flow rate is lower than 1.0 litres/min

=0, function off

Range: 0.01-655.35

Default value:1.0

4.       4.00000: Total amount alarming value. When =0, Function off.

Range: 1-65535

Default value: 0, no display, no total volume alarm output.

5.       5.00000: Alarming time period. When =0, Function off.

Range: 1-65535

Default value: 0, always alarm after triggerred,  when=3, alarm for 3 seconds and then off.

6.       6.00002: Decimal,

Range: 1-3

Default value: 2. the display number will show like 2.35(L/Min)

 

How to calibrate

Use K (constant) to calibrate the flow rate

Frequency = K x Flow rate

If K set on 10, and the flow rate is 4 lts/min

To calibrate to 8 lts/min, change K to 5 (10x4/8)


US508MT has RS485 Communication Function, The protocol is as following

Model: US508MT

Working Voltage: DC 5-12V

Introduction

This series of products follows the MODBUS protocol instructions. Available command codes are 0x03, 0x06, and 0x10. CRC check: low bit first, high bit last.

Example: Use the 03 instruction to read multiple registers

0x010x030x000x000x000x030x050xCB

(Continuously read the values of 3 registers from the 0x0000 register of device No. 0x0001)

To write to a single register use the 06 instruction

0x010x060x000x010x000x010x190xCA

(Write 0x0001 to the 0x0001 register of device No. 0x0001)

Writing to multiple registers uses 10 instructions

0x010x100x000x000x000x020x040x000x000x000x010x320x6F

(Write 0x0000 to the 0x0000 register of device No. 0x0001, and write 0x0001 to the 0x0001 register of device No. 0x0001)

Default Settings:

Baud rate: 9600 (decimal)

MODBUS device address: 1 (decimal)

Flowmeter constant K value: 500 (decimal)

 

To modify baud rate parameters

The default baud rate is 9600, and the reading is 2.

If you modify the baud rate, modify the parameters as follows. (If you enter a value other than the value in the table, the baud rate will be restored to 9600, but when reading, it will be displayed as the modified value is not 2)

Setting

Baud Rate

0

2400

1

4800

2

9600

3

19200

 

 

Register Address

Register Function

0x0000

Baud rate

0x0001

Device address, can be modified to 0x0001-0x00F7 (decimal 1-247)

0x0002

Sensor pulse K factor, the value read and written is 100 times the actual value

0x0003

Permanent cumulative flow higher register, unit, L (maximum 65535)

0x0004

Permanent cumulative flow lower register, unit, L (maximum 65535)

0x0005

Permanent cumulative flow decimal, the reading is 100 times the actual value, unit, L

0x0006

Permanent accumulated flow pulse number(accumulated 1L to return to 0)

0x0007

Temporary accumulated flow higher register, unit, L (maximum 65535)

0x0008

Temporary accumulated flow lower register, unit, L (maximum 65535)

0x0009

Temporary cumulative flow decimal, the reading is 100 times the actual value, unit, L

0x000A

Temporary accumulated flow pulse number (accumulated 1L to return to 0)

0x000B

Real-time flow rate, unit: L/min,

 the reading is 100 times the actual value

0x000C

Real-time(per second) pulse number unit Hz

 

Note:

1. After the address, baud rate, K factor, and permanent cumulative flow are modified, the device needs to be restarted to take effect.

2. Please close the input pipeline before clearing the cumulative flow to prevent water from passing by again after clearing. The value changes after 60 seconds, and the program changes the value again so that the cumulative amount cannot be returned to zero.

3. Reset the temporary cumulative flow after power failure, and record the permanent cumulative flow every 60 seconds. (Flows must exceed 1 litre within 60 seconds to be recorded)

4. When the low bit of the register is full of 65535, it will enter 1 into the higher bit. Example: Assuming a flow sensor with a cumulative flow higher register is 1 and the cumulative flow lower register is 2; then the total flow is 65535 * 1 +2 = 65537L

5. The maximum storage value of each register is 65,535, and the maximum storage data of the permanent cumulative flow of each flow sensor is 65,535 * 65,535 +65,535 = 4,294,917,600 liters of water

6. During the reset operation, the registers for the permanent accumulated traffic need to be reset at the same time, that is, the registers 3/4/5/6 should be reset at the same time. Temporary cumulative traffic reset is also subject to this rule.