The flow rate displayed is in litres per minute.
How to set
Provide 5V DC power to light on.
Open the back cover of the device
Press K1 and K2 together, to enter the menu or parameters
Then press K1 to select the digit
Press K2 to increase the number
Parameters
1. Type=0 frequency counter
2. Dot – Decimal (0-3)
Example: Dot = 1 means one decimal
It is related to K1 (constant)
If Dot = 1, K1 = 10
If Dot = 2, K1 = 100
3. K1 – constant (see point 4); K2 – constant to calibrate flow rate;
For example: the formula of the flow sensor is F=8Q, then K1=1, K2=8;
4. UNT=1, L/min
How to calibrate
Use K2 (constant) to calibrate the flow rate
Frequency = K2 x Flow rate
If K2 set on 10, and the flow rate is 4 lts/min
To calibrate to 8 lts/min, change K2 to 5 (10x4/8)